首页 | 本学科首页   官方微博 | 高级检索  
     


Calibration results obtained with Liulin-4 type dosimeters.
Authors:Ts Dachev  B Tomov  Yu Matviichuk  Pl Dimitrov  J Lemaire  Gh Gregoire  M Cyamukungu  H Schmitz  K Fujitaka  Y Uchihori  H Kitamura  G Reitz  R Beaujean  V Petrov  V Shurshakov  V Benghin  F Spurny
Affiliation:Solar-Terr. Influences Laboratory-BAS, Sofia, Bulgaria. tdachev@bas.bg
Abstract:The Mobile Radiation Exposure Control System's (Liulin-4 type) main purpose is to monitor simultaneously the doses and fluxes at 4 independent places. It can also be used for personnel dosimetry. The system consists of 4 battery-operated 256-channel dosimeters-spectrometers. We describe results obtained during the calibrations of the spectrometers at the Cyclotron facilities of the University of Louvain, Belgium and of the National Institute of Radiological Sciences-STA, Chiba, Japan with protons of energies up to 70 MeV. The angular sensitivities of the devices are studied and compared with Monte-Carlo predictions. We also present the results obtained at the HIMAC accelerator with 500 MeV/u Fe ions and at the CERN high energy radiation reference fields. Records made during airplane flights are shown and compared with the predictions of the CARI-6 model.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏