Calibration results obtained with Liulin-4 type dosimeters. |
| |
Authors: | Ts Dachev B Tomov Yu Matviichuk Pl Dimitrov J Lemaire Gh Gregoire M Cyamukungu H Schmitz K Fujitaka Y Uchihori H Kitamura G Reitz R Beaujean V Petrov V Shurshakov V Benghin F Spurny |
| |
Affiliation: | Solar-Terr. Influences Laboratory-BAS, Sofia, Bulgaria. tdachev@bas.bg |
| |
Abstract: | The Mobile Radiation Exposure Control System's (Liulin-4 type) main purpose is to monitor simultaneously the doses and fluxes at 4 independent places. It can also be used for personnel dosimetry. The system consists of 4 battery-operated 256-channel dosimeters-spectrometers. We describe results obtained during the calibrations of the spectrometers at the Cyclotron facilities of the University of Louvain, Belgium and of the National Institute of Radiological Sciences-STA, Chiba, Japan with protons of energies up to 70 MeV. The angular sensitivities of the devices are studied and compared with Monte-Carlo predictions. We also present the results obtained at the HIMAC accelerator with 500 MeV/u Fe ions and at the CERN high energy radiation reference fields. Records made during airplane flights are shown and compared with the predictions of the CARI-6 model. |
| |