压力和速度对氧化锆分析仪测量值的影响
引用本文:赵科,段翠九,谭力,矫维红,吕清刚.压力和速度对氧化锆分析仪测量值的影响[J].航空计测技术,2011(6):27-30.
作者姓名:赵科  段翠九  谭力  矫维红  吕清刚
作者单位:[1]中国科学院工程热物理研究所,北京100190 [2]中国科学院研究生院,北京100190
基金项目:国家自然科学基金资助项目(50906084); 中国博士后科学基金资助项目(20090450578); 中国科学院知识创新工程重要方向项目(KGCX2-YX-399+3)
摘    要:使用氧化锆分析仪测量气体中氧浓度时发现,压力和速度可能对测量值有影响。本文组建实验系统并进行实验研究,用氧气和空气混合制成高氧气浓度气体,用氧气和二氧化碳混合制成低氧气浓度气体,分别研究了压力和速度对不同氧气浓度下氧化锆分析仪测量值的影响。结果表明,氧化锆分析仪在直接插入式安装时,存在最大使用压力;在最大使用压力以下测量时,气体压力对测量值无影响。压力超过最大使用压力时,高氧气浓度下,随着压力升高,测量值减小;低氧气浓度下,随着压力升高,测量值增加。气流速度对分析仪的测量值无影响。

关 键 词:氧化锆分析仪  压力  速度  氧气浓度

Effect of Pressure and Velocity on Measurements of Zirconia Oxygen Analyzer
ZHAO Ke,DUAN Cuijiu,TAN Li,JIAO Weihong,LV Qinggang.Effect of Pressure and Velocity on Measurements of Zirconia Oxygen Analyzer[J].Aviation Metrology & Measurement Technology,2011(6):27-30.
Authors:ZHAO Ke  DUAN Cuijiu  TAN Li  JIAO Weihong  LV Qinggang
Institution:1.Institute of Engineering Thermophysics,Chinese Academy of Sciences,Beijing 100190,China; 2.Graduate School of the Chinese Academy of Sciences,Beijing 100190,China)
Abstract:The gas pressure and velocity maybe affect the measurements when using the zirconia oxygen analyzer to measure the oxygen concentration.In this paper,an experimental study was made to investigate the effect of the pressure and velocity on measurements in different oxygen concentration gases.High oxygen concentration gas was achieved by blending air and oxygen.Low oxygen concentration gas was obtained by mixing carbon dioxide and oxygen.The results show that the gas pressure must be lower than the maxmum application pressure(MAP) when the analyzers were directly installed into the system.Within MAP,the gas pressure has no effect on the measurements.For pressure exceeding the MAP in high oxygen concentration,the measurement result of oxygen concentartion decreases with the increase of pressure.However,the rules are inversed in low oxygen concentration.The gas velocity has no effect on the measurements of the analyzer.
Keywords:zirconia oxygen analyzer  pressure  velocity  oxygen concentration