科氏质量流量计非线性振动特性研究(英文)
引用本文:郑德智,王帅,樊尚春.科氏质量流量计非线性振动特性研究(英文)[J].中国航空学报,2009,22(2):198-205.
作者姓名:郑德智  王帅  樊尚春
作者单位:Zheng Dezhi,Wang Shuai(School of Instrument Science and Opto-electronics Engineering,Beijing University of Aeronautics and Astronautics,Beijing 100191,China;Key Laboratory of Precision Opto-mechatronics Techonology,Ministry of Education of the People's Republic of China,Beijing);Fan Shangchun(Key Laboratory of Precision Opto-mechatronics Techonology,Ministry of Education of the People's Republic of China,Beijing 100191,China) 
基金项目:National High-tech Research and Development Pro-gram (2008AA042207)
摘    要:科氏质量流量计(Coriolis mass flowmeter,CMF)当测量管振动幅度较小时,一般被认为是线性系统,但CMF输出信号频谱中具有倍频成分。结合CMF传感器的数学模型,在适当简化测量管边界条件的情况下,建立了考虑非线性因素和外部激励的直管型CMF测量管运动方程,并运用Galerkin方法对运动方程进行离散处理,由离散后的结果可以看出,系统含有多组非线性影响因素成分。针对测量管的闭环谐振和标准正弦信号受迫振动,分别采用激光测振仪对其振动状态进行测量,测量结果均含有非线性因素成分,证实了非线性分析的准确性;针对测量管的不同振动幅度进行了测量精度实验,验证了测量管振动幅度越大非线性影响越严重。理论分析结果和实验结果对CMF性能的进一步提高具有指导意义。

收稿时间:4 September 2008

Nonlinear Vibration Characteristics of Coriolis Mass Flowmeter
Zheng Dezhi,Wang Shuai,Fan Shangchun.Nonlinear Vibration Characteristics of Coriolis Mass Flowmeter[J].Chinese Journal of Aeronautics,2009,22(2):198-205.
Authors:Zheng Dezhi  Wang Shuai  Fan Shangchun
Institution:1. School of Instrument Science and Opto-electronics Engineering, Beijing University of Aeronautics and Astronautics, Beijing 100191, China;2. Key Laboratory of Precision Opto-mechatronics Techonology, Ministry of Education of the People''s Republic of China, Beijing 100191, China
Abstract:Coriolis mass flowmeter (CMF) is generally regarded as a linear system when the vibration amplitude of the measuring tube is relatively small, but the spectrum of its output signals still show existence of some multiple frequency components. Taking into consideration the nonlinear factors and external excitation, the motion equations of the CMF straight measuring tube are estab- lished by properly simplifying the boundary conditions of the measuring tube and integrating the mathematical model of the CMF sen...
Keywords:flowmeters  Coriolis  closed loop control systems  nonlinear systems  vibrations