首页 | 本学科首页   官方微博 | 高级检索  
     


Preliminary calibration results of VEGA 1 and 2 SIGMA-3 gas chromatograph
Authors:L.M. Mukhin, D.F. Nenarokov, N.V. Porschnev, V.B. Bondarev, B.G. Gelman, G. Israe&#x  l, F. Raulin, J. Runavot,R. Thomas
Affiliation:

*Space Research Institute, Moscow, U.S.S.R.

French Space Agency, CNES, 31000 Toulouse, France

LPCE, CNRS 45000 Orle´ans, France

**Service d'Ae´ronomie, CNRS, 91370 Verrie`res-le-Buisson, France

***Universite´Paris Val de Marne, 94000 Cre´teil, France

Abstract:SIGMA - 3 gas chromatograph on board VEGA 1 and 2 landing probes has been operated successfully in the 60 - 50 km altitude range, providing several in - situ chemical analysis of the gas and the aerosols of Venus cloud layers. Post flight calibration required to derive atmospheric abundancies from gas chromatograms were carried out using the SIGMA - 3 spare model. A Venus atmospheric aerosol simulation chamber was used in which sulfuric acid droplets were generated. Preliminary results of these calibration experiments indicate that the concentration of sulfuric acid in the upper part of the clouds ( 60 to 55 km) is about 1 mg/m3 and suggest that an additional constituant must be present in noticeable amount in the aerosols. From these experiments the mixing ratio upper limits of SO2 is 100 ppmV and of H2S and COS is few 10 ppmV.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏