首页 | 本学科首页   官方微博 | 高级检索  
     检索      

用于EMI测量接收机脉冲响应校准的标准脉冲信号产生方法
引用本文:刘爽,马红梅,姜河,龚鹏伟,杨春涛,谢文,谌贝,袁艳.用于EMI测量接收机脉冲响应校准的标准脉冲信号产生方法[J].宇航计测技术,2022,42(3):7-13.
作者姓名:刘爽  马红梅  姜河  龚鹏伟  杨春涛  谢文  谌贝  袁艳
作者单位:1.北京无线电计量测试研究所,北京100039;2.计量与校准技术重点实验室,北京 100039
摘    要:EMI测量接收机作为电磁兼容测试中的关键仪器应用越来越广泛,但目前该类仪器的脉冲响应校准所用的标准源完全依赖于进口,国内尚无可替代的标准脉冲信号产生器。介绍一种用于EMI测量接收机脉冲响应校准的标准脉冲信号产生方法,首先建立标准脉冲信号的参数模型并对其进行数值分析和仿真,验证其符合CISPR 16-1-1标准的要求,之后搭建实验系统,利用标准脉冲信号生成软件控制任意脉冲信号发生器输出标准脉冲信号并对其进行补偿修正,获得符合CISPR 16-1-1标准要求的标准脉冲信号,该信号可用于EMI测量接收机脉冲响应的校准。

关 键 词:EMI测量接收机  脉冲响应校准  标准脉冲信号  参数模型  数值分析  信号预失真  

Standard Pulse Signal Generation Method for Impulse Response Calibration of EMI Measurement Receiver
LIU Shuang,MA Hong-mei,JIANG He,GONG Peng-wei,YANG Chun-tao,XIE Wen,CHEN Bei,YUAN Yan.Standard Pulse Signal Generation Method for Impulse Response Calibration of EMI Measurement Receiver[J].Journal of Astronautic Metrology and Measurement,2022,42(3):7-13.
Authors:LIU Shuang  MA Hong-mei  JIANG He  GONG Peng-wei  YANG Chun-tao  XIE Wen  CHEN Bei  YUAN Yan
Institution:1.Beijing Institute of Radio Metrology and Measurement,Beijing 100039,China;2.Science and Technology on Metrology and Calibration Laboratory,Beijing 100039,China
Abstract:As a key instrument in electromagnetic compatibility testing,EMI measurement receiver has been used more and more widely.However,the standard source used for pulse response calibration of this kind of instrument completely depends on imports,and there is no substitute for the standard pulse signal generator in the domestic market.This article introduces a standard pulse signal parameter analysis method for impulse response calibration of EMI measurement receiver.First,we establish the parametric model of the standard pulse signal and perform numerical analysis and simulation to verify whether it meets the requirements of the standard CISPR 16-1-1.After that we build the experimental system,and develop the standard pulse signal generation software which controls the arbitrary pulse signal generator to output the standard pulse signal and compensates it,thus the standard pulse signal which meets the requirements of CISPR 16-1-1 is obtained.The signal can be used to calibrate the pulse response of EMI measuring receiver.
Keywords:EMI measurement receiver  Pulse response calibration  Standard pulse signal  Parametric model  Numerical analysis  Signal predistortion  
点击此处可从《宇航计测技术》浏览原始摘要信息
点击此处可从《宇航计测技术》下载免费的PDF全文