首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The Search-Coil Magnetometer for MMS
Authors:Email author" target="_blank">O?Le?ContelEmail author  P?Leroy  A?Roux  C?Coillot  D?Alison  A?Bouabdellah  L?Mirioni  L?Meslier  A?Galic  M?C?Vassal  R?B?Torbert  J?Needell  D?Rau  I?Dors  R?E?Ergun  J?Westfall  D?Summers  J?Wallace  W?Magnes  A?Valavanoglou  G?Olsson  M?Chutter  J?Macri  S?Myers  S?Turco  J?Nolin  D?Bodet  K?Rowe  M?Tanguy  B?de?la?Porte
Institution:1.Laboratoire de Physique des Plasmas,CNRS/Ecole Polytechnique/Sorbonne Universités UPMC/Paris-Sud 11,Palaiseau,France;2.Laboratoire Atmosphère, Milieux, Observations Spatiales,CNRS/UVSQ/UPMC,Guyancourt,France;3.Nexeya Conseil & Formation,Toulouse,France;4.3D PLUS,Buc,France;5.University of New Hampshire,Durham,USA;6.University of Colorado,Boulder,USA;7.Space Research Institute,Austrian Academy of Sciences,Graz,Austria;8.Royal Institute of Technology,Stockholm,Sweden;9.University of California,Los Angeles,USA
Abstract:The tri-axial search-coil magnetometer (SCM) belongs to the FIELDS instrumentation suite on the Magnetospheric Multiscale (MMS) mission (Torbert et al. in Space Sci. Rev. (2014), this issue). It provides the three magnetic components of the waves from 1 Hz to 6 kHz in particular in the key regions of the Earth’s magnetosphere namely the subsolar region and the magnetotail. Magnetospheric plasmas being collisionless, such a measurement is crucial as the electromagnetic waves are thought to provide a way to ensure the conversion from magnetic to thermal and kinetic energies allowing local or global reconfigurations of the Earth’s magnetic field. The analog waveforms provided by the SCM are digitized and processed inside the digital signal processor (DSP), within the Central Electronics Box (CEB), together with the electric field data provided by the spin-plane double probe (SDP) and the axial double probe (ADP). On-board calibration signal provided by DSP allows the verification of the SCM transfer function once per orbit. Magnetic waveforms and on-board spectra computed by DSP are available at different time resolution depending on the selected mode. The SCM design is described in details as well as the different steps of the ground and in-flight calibrations.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号