Affiliation: | * IRDI/DMECN/DMG/SEM Laboratoire d'Etude de la Solidification, 38041, Grenoble Cedex, France ** IRDI/DMECN/DMG/SDTT Section des Ateliers de Prototypes et de Techniques Avancées, Centre d'Etudes Nucléaires, 85 X, 38041, Grenoble Cedex, France |
Abstract: | The drop tube which will be available in the “Centre d'Etudes Nucléaires de Grenoble” is described. Its main features are the following: - Dimensions : Drop height : 47.1 m Drop time : 3.1 s Tube inside diameter : 0.2 m - Experimental atmosphere : 1 Ultra-vacuum : 10−6 to 10−7 Pa - Residual gravity level : 10−8 to 10−9 g according to the vacuum level and drop diameter. This facility is unique insofar as it enables experiments to be performed under ultra-vacuum conditions which, by delaying the formation of surface oxides, should contribute to improving maximum undercooling values. The techniques used for obtaining small metallic drops (0.5 to 3 mm) are described. The availability of this instrument for the scientific community is also foreseen by the french sponsoring organizations (CEA, CNES, CNRS) ; some practicle informations will be given to potential experimenters. |