A.S. Laxmiprasad V.L.N. Sridhar Raja Surya Menon Adwaita Goswami M.V.H. Rao K.A. Lohar 《Advances in Space Research (includes Cospar's Information Bulletin, Space Research Today)》2013
A miniaturized in situ laser induced breakdown spectroscope-LIBS is one of the two lunar rover payloads to be flown in India’s next lunar mission Chandrayaan-2, with an objective to carry-out a precise qualitative and quantitative elemental analyses of lunar regolith at the proximity of the landing region. As per the imposed mission constraints and the executed design optimization studies, a compact and light-weight LIBS prototype model is developed at our premises. This paper mainly concerns with the estimation of theoretical aspects; especially on evaluation of elemental ablation parameters and signal-to-noise ratio (SNR) calculations for the designed instrument. Theoretical estimations and simulations yielded an incident laser power density of the order of 5 × 1010 W/cm2 on the target surface at a defined lens-to-surface distance (LTSD) of 200 mm and revealed an SNR > 100 for most of the elements under consideration. This paper also addresses the impact of LTSD variation on detection capability. The estimation of plasma-temperatures was carried out utilizing the emission spectra obtained under high vacuum environments employing the LIBS laboratory model. Experimental investigations and the performed theoretical estimations asserted the successful operation of the configured LIBS instrument for in situ elemental analyses on lunar surface. 相似文献